Skip to content

Possible inaccuracy in attenuation length for BC-420 #1

Description

@stealex

Hello!

First of all, congratulations for the nice work!

I am by no means an expert on all of this so whatever is below might be a misunderstanding on my part. In that case, I apologize and I would be grateful for any clarification.

I've been trying to find the optical properties of BC-420 and I've stumbled upon this repository. While trying to get everything to work in my application, I started cross-checking the numerical values to make sure I understand everything properly. I am comparing entries of opsc-106 with the data I could find here
https://luxiumsolutions.com/radiation-detection-scintillators/plastic-scintillators/fast-timing-bc-418-bc-420-bc-422-bc-422q

Please correct me if I'm wrong but: the absorption length (light attenuation length) is stated in the datasheet as 140 cm. The repository uses 120 cm (https://github.com/mkandemirr/SSLG4/blob/main/sslg4/data/oscnt/opsc-106/absLength.txt).

Activity

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment